CASE REPORT |
https://doi.org/10.5005/jp-journals-10003-1365 |
Noise through Bone Conduction—Differential Diagnosis for the Type of Hearing Loss
1,2Department of Speech and Hearing, JSS Institute of Speech and Hearing, Mysuru, Karnataka, India
Corresponding Author: Sindhu Parthasarathy, Department of Speech and Hearing, JSS Institute of Speech and Hearing, Mysuru, Karnataka, India, Phone: +91 08212548229, e-mail: sindhunp11@gmail.com
How to cite this article: Parthasarathy S, Shetty HN. Noise through Bone Conduction—Differential Diagnosis for the Type of Hearing Loss. Int J Otorhinolaryngol Clin 2021;13(1):26–28.
Source of support: Nil
Conflict of interest: None
ABSTRACT
Aim: The study aims to assess the threshold shift in sensorineural acuity level (SAL) test and bone conduction speech recognition threshold (BCSRT) in mixed and sensorineural hearing loss.
Background: The diagnosis of moderate mixed hearing loss and sensorineural hearing loss is unclear when the air-bone gap (ABG) is at 15 to 20 dB. The study was designed to arrive at the accurate diagnosis of moderate mixed hearing loss and sensorineural hearing loss.
Case description: The study includes five participants diagnosed with bilateral normal hearing (n = 1), bilateral conductive hearing loss (n = 1), bilateral sensorineural hearing loss (n = 1), and those with bilateral moderate loss with an ABG of 15 to 20 dB (n = 2). Conventional pure tone audiometry, speech audiometry, immittance, reflexes, and distortion product otoacoustic emission were administered.
Conclusion: The BCSRT was almost the same in confirmed normal hearing and conductive hearing loss participants because their BC sensitivity was within the normal limit. BCSRT was high in individuals with sensorineural hearing loss and mixed hearing loss. In addition, the amount of threshold shift in SAL was taken into consideration for the diagnosis.
Clinical significance: The amount of threshold shift in the SAL test with accompanying BCSRT concludes that the presence and absence of infection in the middle ear thereby lead to an accurate diagnosis.
Keywords: Bone conduction speech recognition threshold, Mixed hearing loss, Sensorineural acuity level.
BACKGROUND
Many a time diagnoses of moderate mixed hearing loss and sensorineural hearing loss (SNHL) are confusing if a patient’s air–bone conduction gap (ABG) is at 15–20 dB with A-type tympanogram. The possible reason is the variability of pure tone thresholds which accounts for 5–10 dB.1 Most often to lessen confusion, other diagnostic tests such as otoacoustic emission and reflex thresholds are administered. But, these test results may show absence due to hearing loss. Even then confusion is still being present but administering these tests to the test battery merely adds up unnecessary clinical expenses and increases in clinical time. Considering cost and time factors, it is imperative to use a sensorineural acuity test. In the SAL test, narrowband noise is presented through the bone conduction at the forehead. The difference between masked and unmasked thresholds is calculated. The shift in threshold after introducing a noise gives an idea of the origin of the problem in the auditory pathway. It is a well-established fact that bone conduction is a test to assess the acuity of a sensorineural component. Thus, it is assumed that sensorineural acuity level may accurately diagnose patients with having either SNHL or mixed hearing loss, especially for patients who have an ABG of 15–20 with A tympanogram. It is hypothesized that if a patient is suffering from SNHL, then the shift between the unmasked and masked threshold would be less than in a patient suffering from mixed hearing loss. In addition, to validate SAL, the bone conduction speech recognition test (BCSRT) was also administered. The following objective is utilized to assess the threshold shift in SAL (at 500 Hz, 1 and 2 kHz) and BCSRT from the participants of the study.
CASE DESCRIPTION
A total of five participants were involved to determine the threshold shift in SAL and BCSRT from the study participants. Conventional pure tone audiometry,2 speech audiometry,3 immittance, and reflexes4 were administered on these five participants. In addition, if required, distortion product otoacoustic emission (DPOAE) was administered. The results of audiological evaluation revealed that the participants had confirmed bilateral normal hearing sensitivity (hereinafter AB), bilateral moderate conductive hearing loss (hereinafter AC), and bilateral moderate sensorineural hearing loss (hereinafter AD) (Fig. 1). However, the other two participants had bilateral moderate loss with ABG of 15 to 20 dB with A-type tympanogram, absent reflexes, and absent DPOAEs in them. These audiological findings in them were unable to be accurately diagnosed. Thus, SAL and BCSRT were performed. The complete audiological profile of all five participants is represented in Table 1.
Particulars | AB | AC | AD | AE | AF |
---|---|---|---|---|---|
Age (years) | 25 | 40 | 55 | 49 | 53 |
Complaints | — | Blocking sensation | Reduced hearing | Blocking sensation and reduced hearing | Unable speech in noise and reduced hearing |
PTA (dB HL) | 10 | 55 | 50 | 53.3 | 56.3 |
Average ABG (dB) (from 0.25 to 2 kHz) | — | 40 | 5 | 20 | 20 |
SRT dB | 10 | 45 | 45 | 50 | 45 |
SIS (%) | 100 | 100 | 96 | 96 | 96 |
Tympanogram | A | B | A | A | A |
Reflexes | Ipsi- and contra-present | Ipsi- and contra-absent | Ipsi- and contra-elevated | Ipsi- and contra-absent | Ipsi- and contra-absent |
DPOAEs | Present | Absent | Absent | Absent | Absent |
Provisional diagnosis | Hearing sensitivity within the normal limit | Conductive hearing loss | SNHL | ?? | ?? |
Sensorineural Acuity Level Test
The procedure given by Jerger and Tillman5 was adopted to assess the sensorineural acuity level. Initially, air conduction pure tone threshold for frequencies 0.25 to 2 kHz (in octaves) was obtained through bracketing method (−10 and +5 dB).6 A bone vibrator was positioned at the forehead and a circumaural headphone was placed on each participant’s ears. A narrowband masking noise corresponding to a pure tone was delivered through a calibrated bone vibrator at a maximum level of 55 dB. Masked air conduction pure tone threshold for frequencies 0.25 to 2 kHz in octaves was obtained. A threshold shift was calculated by taking the difference between masked and unmasked thresholds (Fig. 2).
BCSRT
The bone vibrator was positioned at the test-ear mastoid. Three spondees were presented at +12 dB above the best BC. If two or three spondees were repeated correctly out of three, then the intensity was reduced by 5 dB. If only one spondee or unable to repeat it correctly out of three, then the intensity is increased by 10 dB. With this bracketing method, a low-intensity level at which at least 2/3 responses were obtained was considered as BCSRT.To derive genuine BCSRT in the target test ear, a masking noise was presented to the contralateral ear through an insert receiver. The level of noise delivered was the AC threshold of the contralateral ear and its ABG (Fig. 3).
DISCUSSION
The threshold shift in the SAL test for each frequency and BCSRT from study participants was descriptively analyzed. The results of the SAL test for 1 kHz and BCSRT from the study participants are depicted in Figure 1. The markings of 1, 2, and 3 in Figure 1 represent an unmasked threshold, masked threshold, and a threshold shift, respectively, for 1 kHz from the SAL test for each participant.
The BCSRT was almost the same in confirmed normal hearing (AB) and conductive hearing loss (AC) participants because their BC sensitivity was within the normal limit. However, BCSRT was obtained at 45 dB in confirmed cases of SNHL (AD). In the case of SNHL, the cochlea is affected and the destiny of the BC pathway at the peripheral level is the cochlea. In SNHL, the cochlea is damaged; thus, the level required for BCSRT was high. The BCSRTfor AE and AF were 25 and 45 dB, respectively. The results BCSRT gives are the notion that in both cases the cochlea is affected but a relatively higher damage might be in the case of AF than AE which was reflected in the BCSRT test.
CONCLUSION
The amount of threshold shift in the SAL test with accompanying BCSRT concludes the presence and absence of infection in the middle ear, thereby leading to an accurate diagnosis.
CLINICAL SIGNIFICANCE
The results of the SAL test on these dilemma cases of the present study throw light on accurate diagnosis. It gives a sign of infection if present in the middle ear to the otolaryngologist for further management.
ACKNOWLEDGMENT
The authors extend their sincere gratitude to all the participants of the study.
ORCID
Sindhu Parthasarathy https://orcid.org/0000-0002-9783-986X
Hemanth N Shetty https://orcid.org/0000-0002-5161-1368
REFERENCES
1. Schmuziger N, Probst R, Smurzynski J. Test-retest reliability of pure-tone thresholds from 0.5 to 16 kHz using Sennheiser HDA 200 and Etymotic Research ER-2 earphones. Ear Hear 2004;25(2):127–132. DOI: 10.1097/01.aud.0000120361.87401.c8.
2. American Speech-Language-Hearing Association. Guidelines for manual pure- tone threshold audiometry [Guidelines]; 2005. Available from: www.asha.org/policy.
3. American Speech, Language and Hearing. Determining threshold level for speech; 1988. p. 85–89.
4. American Speech-Language-Hearing Association. Guidelines for acoustic immittance screening of middle ear function. Asha 1979;21(288):563–569.
5. Jerger J, Tillman T. A new method for the clinical determination of sensorineural acuity level (SAL). AMA Arch Otolaryngol 1960;71(6):948–955. DOI: 10.1001/archotol.1960.03770060060008.
6. Carhart R, Jerger JF. Preferred method for clinical determination of pure-tone thresholds. J Speech Hear Disord 1959;24(4):330–345. DOI: 10.1044/jshd.2404.330.
________________________
© The Author(s). 2021 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by-nc/4.0/), which permits unrestricted use, distribution, and non-commercial reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.